Home arrow MIP arrow Detector Comparisons

Geoprobe® Systems has developed Detector Comparisons that are completed by using the "Response Test Procedure" located in the MIP SOP Technical Bulletin MK3010. With this procedure, the probe is immersed for a period of 45 seconds into a cylinder of water containing a dissolved contaminant. This allows the operator to determine if the system is functioning properly, if the membrane is in working order, and if the contaminant of concern can be detected by the current detector scheme.

Printable version of the above detector comparisons.

Membrane Response vs. Depth

The chart below shows the results of response testing an MIP system while performing a series of logs. This data was collected by Geoprobe® Systems while working on a site contaminated with TCE. A response test was conducted before and after each MIP log. The boxes indicate the current day ending response test and the next day beginning response test. A new membrane was inserted after the first hole on the third day, due to a membrane rupturing on removal of the tool from the subsurface.

Response test data for each MIP project should be plotted in this manner. Besides being a test of the membrane, and indeed the entire MIP system, response test data allows the user to normalize data to a common level.

MIP Repeatability Chart

Membrane Response to 500 µg/L Standard of TCE

10/29/2003 through 11/01/2003
Response test of TCE run before and after each logged MIP location.
Response test run per Geoprobe® Systems SOP.

Depth achieved
before Standard
in Feet
Baseline to
Peak Δ (mV)
Notes
0 111 Day 1
17 89  
36 76  
54 85
54 91 Day 2
72 103  
87 80  
104 100  
104 110 Day 3
122 202 New Membrane (ruptured on pull)
140 198  
175 130  
175 114 Day 4
193 84  
213 70